Keratoconus and Presbyopia: Wavefront-Guided EDOF Scleral Lens

Case Report
GSLS 2022

Wavefront-Guided EDOF Optics in a Presbyopic Patient With Keratoconus

  • John D. Gelles, OD · Cornea and Laser Eye Institute–Hersh Vision Group, CLEI Center for Keratoconus, Teaneck, NJ; Rutgers New Jersey Medical School, Newark, NJ
  • Nicolas Brown, BS · OVITZ Corporation, Rochester, NY
  • Andres Guevara, PhD · OVITZ Corporation, Rochester, NY
  • Sydney Krisa, BS · CLEI, Teaneck, NJ
  • Daniel Chung, BS · CLEI, Teaneck, NJ
  • Steven A. Greenstein, MD · CLEI, Teaneck, NJ; Rutgers New Jersey Medical School, Newark, NJ
  • Peter S. Hersh, MD · CLEI, Teaneck, NJ; Rutgers New Jersey Medical School, Newark, NJ

A 49-year-old presbyopic patient with keratoconus compared traditional, wavefront-guided and wavefront-guided EDOF scleral optics. The EDOF lens reduced HORMS versus traditional optics in the right eye and improved acuity at all distances, reaching 20/20 near in each eye.

At a glance

Patient

49-year-old woman with keratoconus and presbyopia

Problem

Near acuity of 20/30 in each eye with a distance-corrected traditional scleral lens

Lens

Traditional optics, wavefront-guided optics and wavefront-guided EDOF optics designed on the xwave, each worn 2 weeks

Result

Near 20/30 → 20/20 OD and OS; distance 20/20 → 20/15+ OU

Results

MeasureTraditionalWavefront-guidedWavefront-guided EDOF
Distance VA OD / OS / OU20/20 / 20/20 / 20/2020/15 / 20/15 / 20/15+20/15 / 20/20 / 20/15+
Near VA OD / OS / OU20/30 / 20/30 / 20/2520/30 / 20/25− / 20/2520/20 / 20/20 / 20/20
HORMS OD (7.4 mm)1.34 µm0.49 µm0.63 µm
HORMS OS (6.3 mm)0.61 µm0.23 µm0.62 µm

Total higher-order RMS by optic

Traditional
Wavefront-guided
Wavefront-guided EDOF
OD (7.4 mm)
1.34 µm
0.49 µm
0.63 µm
OS (6.3 mm)
0.61 µm
0.23 µm
0.62 µm

Relative to traditional optics, HORMS fell 63% OD and 62% OS with wavefront-guided optics, and 52% OD and 0% OS with the EDOF optic. The authors note the rise from wavefront-guided to EDOF is expected given the nature of EDOF correction. Data were collected after 2 weeks of wear, with at least 3 hours of wear before examination.

Clinical takeaways from the authors

  1. Presbyopia and keratoconus can be addressed together. A deliberately designed wavefront-guided EDOF optic improved distance and near performance in an irregular cornea.
  2. Judge EDOF by function, not RMS. HORMS rose from the single-vision to the EDOF optic by design, while near acuity improved to 20/20.

Limitations

Single case; the authors call for larger prospective studies.

Disclosures

John D. Gelles, OD, has received research support, devices or honoraria from OVITZ. Andres Guevara, PhD, and Nicolas Brown are employed by OVITZ. Other authors report no relevant financial disclosures.

Have patients like this?

OVITZ is built for complex eyes. Every wavefront-guided optic is custom-designed by the OVITZ optical team from your xwave measurements, and we review complex cases with you, including post-crosslinking, CAIRS, post-PK, post-refractive surgery and highly irregular corneas. OVITZ optics work with most major scleral lens brands.

Related

Need Reading Glasses With Your Scleral Lenses? →

For your patients: a plain-language guide you can share

Clinical information notice: This page summarizes a clinical poster for eye care professionals. Individual outcomes vary. It is not a diagnosis or a substitute for evaluation by a qualified eye care professional.